Withstanding Voltage/Insulation
Tester
Back to
index
|
FEATURES: |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
| SPECIFICATIONS | ||
|
AC HI-POT |
Voltage |
0 ~ 2.5kV, 0 ~ 5kV 2 ranges |
| AC Line waveform | ||
| Voltage Regulation | 20% | |
| Max. Current | 100mA | |
| |
||
| Reference Value | Setting | 0.5, 1, 2, 4, 8, 10, 100mA 7 ranges, combination of from 0.5 ~ 25.5mA ranges |
| Low Reference | From 0 to one-half of reference value continuously variable | |
| |
||
| Meter | 2.5kV, 5kV two ranges, ±3% of full scale | |
| |
||
| Current Monitor Meter | ±5% of full scale | |
| |
||
| Insulation Resistance | DC Voltage | 500V/1000V |
| Resistance Range | 1 ~ 2000MΩ | |
| Resistance Accuracy | 5% ~ 10% depends on range | |
| |
||
| Judgment | Description | Window comparator system |
| Accuracy | ±10% ~ ±20% depends on range | |
| |
||
| Timer | 0.5S ~ 10MIN | |
| |
||
| General | Type of Tests | AUTO W-I, AUTO I-W, MANUAL-W, MANUAL-I |
| Remote Control | Test, Reset and Timer On-Off control | |
| Output Signal | Test ON, GOOD, NG, and Read signal (The rating of the contact signal are 110VAC 1A or 30VDC 1A) | |
| AC Outlet | Use only Buzzer and warning light, the output voltage depend on the line voltage and the current is 1A | |
| Test Mode | Double action, good hold, momentary, NG alarm | |
| |
||
| Power Source | AC 120V±10%, 50/60Hz (AC 220V/240V order made) | |
| |
||
| Accessories |
GHT-105 x 1, Instruction manual x 1, Power cord x 1 (GHT-106R for option) |
|
| |
||
| Dimensions & Weight | 430(W) x 223(H) x 510(D) mm, Approx. 28kg | |